TES-1315/1316 K.J.E.T.R.S.N. Data-Logger Thermometer

Features:

TES-1315/1316, TES-1315, TES-1316, K.J.E.T.R.S.N., Data-Logger, Thermometer
•Highly accurate thermometer with thermocouple K, J, E, T, R, S, N type.
•Dual input function T1/T2 temperature display. (1316)
•7500 DataLogging capacity.
•Warning beeper with HI/LO setting.
•MAX/MIN with TIME, MAX-MIN AVG, HOLD function.
•ºC / ºF / K User-Selective.
•Back-Light LCD display.

Specifications:  
Measuring Range K - Type : -200ºC to +1372ºC (-328ºF to +2501ºF)
J - Type : -210ºC to +1200ºC (-346ºF to +2192ºF)
E - Type : -150ºC to +1000ºC (-238ºF to +1832ºF)
T - Type : -250ºC to +400ºC (-418ºF to +752ºF)
R and S - Type : 0ºC to +1767ºC (+32ºF to +3212ºF)
N - Type : -200ºC to +1300ºC (-328ºF to +2372ºF)
Display Resolution 0.1ºC / ºF / K < 2000º
1.0ºC / ºF / K ≤2000º
Measurement Accuracy
(According to ITS-90)
K, J, E, T, and N-Type : +/-[0.05% of reading +0.3ºC (0.6ºF)]
[below -100ºC (-148ºF) : add 0.15 % of reading for K-, J-, E-, and N-Type ; and 0.45% of reading for T-Type]
R-and S-Type : +/-[0.05% of reading + 0.4ºC (0.8ºF)]
Temperature Coefficient 0.01% of reading + 0.03ºC per ºC (0.05 ºF per ºF) outside the specified +18ºC to 28ºC (+64ºF to +82ºF) range
[LCD ranging : below -100ºC (-148ºF) : add 0.04% of reading for K-, J-, E, and N- Type; and 0.08% of reading for T- Type]
Maximum Differential
Common Mode Voltage
1V (Maximum voltage difference between T1 and T2)
Input Protection 30V maximum input voltage on any
Combination of input Connector
Sampling Rate 1 time/sec
Operating/Storage Conditions 0ºC ~ 40ºC < 80% RH ,
-10ºC ~ 60ºC < 70%RH
Power Source 6 pcs size AAA Battery
Battery Life Approx. 300hours. (not including backlight and beeper)
Size 152(L) x 72(W) x 37(H)mm
Weight Approx. 300g
Accessories 1. Instruction manual, Battery
2. Software for WINDOWS, RS-232 Cable, 9 pin to 25 pin Gender changer.