Enforcer Calibration Accessory from Zellweger Analytics
The
Enforcer provides easy and cost-effective 'bump testing' for
the Impact range of multi gas detectors.
Regulations require regular testing of gas monitoring equipment to ensure that the instrument will work as and when required. The Enforcer allows you to easily comply with regulations when using an Impact multi-gas monitor without the need for costly test facilities or third parties. Bump testing should be carried out before use of any portable gas detector. This involves exposing each gas sensor to a known concentration of target gas and ensuring that it responds correctly. If not, the unit must be re-calibrated or serviced to replace those parts not working. Personal gas monitors are normally calibrated every 3 to 6 months. Typically this requires specialist knowledge and equipment that is costly or the work is sub-contracted to a third party. During this time instruments are not available for use and so could delay essential operations or force additional monitors to be purchased. The Enforcer calibrates Impact multi-gas monitors fitted with Oxygen, flammable, Carbon Monoxide and Hydrogen Sulphide sensors in less than two minutes. Volumes and flow rates are minimised to allow approximately 200 calibrations per cylinder of gas making calibration as low in cost as a Bump test. Impact instruments when used with the Enforcer are always
available for use. Calibration results are recorded for
printing at a later date via the optional PC software thereby
ensuring that instrument histories are always available
if required. Convenient to Use
Easy to Operate
Minimize Running Costs
Easy to Maintain
Maintain Records
|